NE214
www.baumer.com 15
9. Programming plan
Lines 11-16 Status for Line 1-6
0 Parameter can be changed
1 Clearing of and/or Input disabled
2 Parameter is skipped
Line 22 Operating modes
0 Progressive preselection
1 Principal preselection
2 Parallel comparison
3 P1 trailing preselection
Line 23 Scaling factor
1.0000 Default
00.0001 Min.
9999.99 Max.
Line 24 Decimal point
0 No decimal point
1 00000.0
2 0000.00
3 000.000
4 00.0000
Line 25 Type of counting
0 Track A possibly UP/DOWN-Signal on track B
1 Subtraction (A-B)
2 Totaling (A+B)
3 Track A 90° B Single evaluation
4 Track A 90° B Double evaluation
5 Track A 90° B Quadruple evaluation
Line 26 Frequency Principal counter Track A
0 10 kHz
1 25 Hz
2 3 Hz
Line 27 Frequency Principal counter Track B
0 10 kHz
1 25 Hz
2 3 Hz
Line 28 Input logic
0 PNP - Operating threshold 11V
1 NPN - Operating threshold 11V
2 PNP - Operating threshold 5V
3 NPN - Operating threshold 5V
4 PNP - Operating threshold 2,5V
5 NPN - Operating threshold 2,5V
Line 29 Reset Principal counter
0 Autom. Reset and ext. static
1 Autom. Reset and ext. diff.
2 External static
3 External diff.
Line 31 Function Signal input 9
0 Stop
1 Hold
2 Keylock
3 Operating time meter ON/OFF
4 Print
Line 33 Output logic for digital outputs
0 Both outputs as Make contacts
1 P1 Break contact, P2 Make contact
2 P1 Make contact, P2 Break contact
3 Both outputs as Break contacts
Line 35 Impulse signal time P1
0.25 Default
0.01 Min.
99.99 Max.
Line 36 Impulse signal time P2
0.25 Default
0.01 Min.
99.99 Max.
Line 37 Take-over of preselections P1,P2
0 Immediately effective
1 With reset
Line 38 Preselection recognition
0 Only with positive range
1 With positive and negative counting range
Comentarios a estos manuales