Bedienungsanleitung_Series09-US_RS232_V1_e.doc 11/11 Baumer Electric AG
26.03.2013 14:43:00/tof Frauenfeld, Switzerland
Teach-in of SDC {0X} {0XA01} Teach-in successful, the
present object position is
stored as SDC
Teach-in of SDE {0Y} {0YB03} Teach-in not successful,
Teach-in range as
defined in the factory
settings is downloaded
Get Configuration {0V} {0VBADC1A1218110
27010000ab53}
Configuration:
B: relative measurement
A: ASCII for periodical
measurements
D: highest sensitivity
C: 4 times averaged
1: Temperature
compensation ON.
A121: P-Code
811027: SW doc no
010000: SW Version
ab: Identification
52: Check sum
Set configuration {0UABAF0}
{0UABAF047} Configuration:
A: absolute
measurement
B: Binary for periodical
measurement
A: lowest sensitivity
F: 32 averagings
0: temperature
compensation OFF
47: Check sum
Write 2 Byte(Identification) {0N01} {0N0123}
01 written to the sensor
Read 2 Byte (Identification) {0O} {0O0124}
01 read from the sensor
Single measurement {0M} {0M11140121}
1: Object within range
1: Echo width big
1401: measured value
Start permanent periodical
data output
{0P} {0P28} This is the first answer,
data follows suit
Error messages (examples)
Error Sent Received Explanation
Wrong address {3M} {0EA82} A = Wrong address
Temperature compensation ON
sent with wrong parameter (3
instead of 1)
{0G3} {0EP97} P = impermissible
parameter
Wrong command {0W} {0EU02} U = unknown command
EOF forgotten {0M {0ET01} T = after SOF before
EOF more than 0.5s
distance between 2
characters
One ‚0’ too many {0M0} {0EF87} F = Framing error =
wrong string length